pg. 24 Rev F 12/08/14
4. Press the ENTER button on the Keypad to turn ON the heater relay
(Q0). The screen to the right will be displayed.
5. Press the ENTER button on the Keypad to turn OFF the heater
relay (Q0). The screen to the right will be displayed.
Warning: If the heater relay (Q0) is left ON, the Extraction Vessel interior will
become very HOT and within five minutes the instrument ETS switch will trip
resulting in the Fault screen shown here.
From this screen press ENTER to get back to the main Diagnostics screen.
6. Press the STOP button on the Keypad to go back to the main
Outputs screen (shown to the right).
7. Press the STOP button on the Keypad to go back to the main
Diagnostics screen (shown to the right).
8. Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more
times to return to your solvent screen (example shown to the
right).
When you exit Diagnostics, heater relay Q0 turns OFF and valves Q1 - Q7 return
to the closed position.
PET Ether
v <ENTER>
Diagnostics
^ <ENTER>
Outputs <ENTER>
Exit <STOP> v
Diagnostics ETS
Fault <ENTER>
Heater (Q0) OFF
ENTER to toggle
Heater (Q0) ON
ENTER to toggle