pg. 18 Rev F 12/08/14
Put the Teflon Insert into the Extraction Vessel and the Extraction Vessel into the instrument.
7.1 Align the Teflon Insert slot with the Extraction
Vessel pin.
The Teflon Insert O-ring and the Extraction Vessel O-ring must be in
place before running the instrument.
The Teflon Insert slot must be aligned with the Extraction Vessel pin
before running the instrument.
7.2 Slide the Extraction Vessel into the instrument ensuring the Extraction Vessel O-ring stays in place.
7.3 Press ENTER on the Keypad.
The Display should now show the screen to the right.
7.4 Pull the Main Handle downward, closing and locking the Extraction Vessel
in place. You will hear a click noise when the Safety Valve Pin locks in
place.
Before running the instrument, verify that the Extraction Vessel is locked in
place by attempting to rotate the Main Handle upward. If it is not locked,
repeat step 7.4 above.
7.5 Verify that the water supply is on.
Press START on the Keypad.
The instrument will run through its automatic operation, showing its
status on the Display. When the instrument has completed its
automatic operation, the Display will show the screen to the right
(showing the solvent that was run last).
At this time, the only light shining on the Display should be the green
power light.
Close Vessel
Water On <START>
Teflon
Insert slot
and
Extraction
Vessel pin